Scale parameters and functions
8.4 Calibration procedure
SIWAREX WP231
Manual, 07/2014, A5E31238908A-02
79
Figure 8-4
Linearizing the scales' characteristic curve
Load
Comment
Load
Digits
L=0
100 kg load cell (2 mV/V) not
loaded
Approx. 0
L0
Mechanical installation on load cell
(dead load)
e.g. 25 kg
Approx. 500 000
L1
Calibration weight 1 placed onto
scale
e.g. 60 kg
Approx. 1 200 000
L2
Calibration weight 2 placed onto
scale
e.g. 80 kg
Approx. 1 650 000
L
max
Nominal weight of load cell
100 kg
Approx. 2 000 000
L
max
+10 %
Rated approx. 10 %
Approx. 110 kg
Approx. 2 200 000