Step 3 - Vessel Shape
1.
Click on Step 3.
2.
Select a predefined vessel shape. (To describe a more complex shape see Linearization
below.)
3.
Click on Apply.
Step 4 - Range
1.
Click on Step 4.
2.
Change units if desired (in meters by default).
3.
Set High and Low Calibration points.
4.
To convert the reading from level to volume enter a value for Maximum Volume.
5.
If a vessel shape with parabolic ends has been selected, set dimensions A and L.
6.
Click on Apply.
Step 5 - Summary
Check parameter settings, and click on Cancel to abort, or Apply to transfer values to the
device.
Linearization
You can use the linearization feature to define a more complex vessel shape and enter up to
32 level breakpoints where the corresponding volume is known. The values corresponding to
100% and 0% levels must be entered. The breakpoints can be ordered from top to bottom,
or the reverse.
Using linearization via the Quick Start wizard
1.
Navigate to Configure/Setup > Wizard > Quick Start.
2.
In Step 1 – Identification, make any desired modifications and click on Apply.
3.
In Step 2 – Application, select a level application and liquid as a material.
4.
In Step 3 – Vessel Shape, choose the vessel shape option Linearization Table.
Содержание SITRANS LR200 7ML5422
Страница 185: ...SITRANS LR200 with mA HART Operating Instructions 05 2019 A5E32337676 AE 185 B 5 4 PFTE rod antenna threaded connection ...
Страница 193: ...SITRANS LR200 with mA HART Operating Instructions 05 2019 A5E32337676 AE 193 ...
Страница 194: ...SITRANS LR200 with mA HART 194 Operating Instructions 05 2019 A5E32337676 AE ...
Страница 212: ...Index SITRANS LR200 with mA HART 212 Operating Instructions 05 2019 A5E32337676 AE ...