52
PBD-51041059
SITRANS LG260 - Operating Instructions
44284-EN-171021
Code
Text message
Cause
Rectification
DevSpec
Diagnosis Bits
F264
Installation/Set-
up error
•
Adjustment not within the vessel
height/measuring range
•
Max. measuring range of the
instrument not sufficient
•
Check for correct mounting and/or
parameter settings
•
Use an instrument with bigger
measuring range
Bit 10
F265
Measurement
function dis-
turbed
•
Sensor no longer carries out a
measurement
•
Operating voltage too low
•
Check operating voltage
•
Carry out a reset
•
Briefly separate operating voltage
Bit 11
F266
Impermissible
operating voltage
•
Wrong operating voltage
•
Check operating voltage
•
Check connection cables
Bit 14
F267
No executable
sensor software
•
Sensor cannot start
•
Exchanging the electronics
•
Send instrument for repair
-
The following table shows the error codes and text messages in the
status message "Function check" and provides information on causes
as well as corrective measures.
Code
Text mes-
sage
Cause
Rectification
DevSpec
State in CMD 48
C700
Simulation ac-
tive
•
A simulation is active
•
Finish simulation
•
Wait for the automatic end
after 60 mins.
"Simulation Active" in "Stand-
ardized Status 0"
The following table shows the error codes and text messages in the
status message "Out of specification" and provides information on
causes as well as corrective measures.
Code
Text mes-
sage
Cause
Rectification
DevSpec State in CMD 48
S600
Impermissi-
ble electronics
temperature
•
Temperature of the process-
ing electronics in the non-
specified section
•
Check ambient temperature
•
Insulate electronics
•
Use instrument with higher
temperature range
Bit 8 of Byte 14 … 24
S601
Overfilling
•
Level echo in the close range
not available
•
Reduce level
•
100 % adjustment: Increase
value
•
Check mounting socket
•
Remove possible interfering
signals in the close range
•
Use coaxial probe
Bit 9 of Byte 14 … 24
Function check
Out of specification