Commissioning
6.3 Application examples
SITRANS LR250 with FOUNDATION FIELDBUS
58
Operating Instructions, 09/2021, A5E32221411-AH
①
Confidence
⑦
Zoom
②
Distance from Low Calibration Point to vertical cross-hair
⑧
Measure
③
Algorithm: t
F
(trueFirst)
⑨
Exit
④
Distance from flange face to target
⑩
Exit icon selected
⑤
Pan left/right - selected
⑪
Exit icon deselected
⑥
Pan up/down
•
Use
UP
or
DOWN
arrow
to scroll to an icon. When an icon is highlighted, that
feature becomes active.
•
To move a cross-hair, press
RIGHT
arrow
to increase the value,
LEFT
arrow
to
decrease.
•
To Zoom into an area, position the intersection of the cross-hairs at the center of that
area, select
Zoom
, and press
RIGHT
arrow.
Press
LEFT
arrow
to Zoom out.
•
To update the profile, select
Measure
and press
RIGHT
arrow.
•
To return to the previous menu, select
Exit
then press
RIGHT
arrow.
6.2.3.5
Device Address
Note
The address can only be changed from a remote master such as NI-FBUS Configurator or
DeltaV. See Addressing in manual Foundation Fieldbus for Level Instruments
(7ML19985MP01) for further details.
Read only. The unique address of the device on the network.
Values
Temporary range during initial commissioning: 248 - 251. Permanent range
after commissioning complete (written to non-volatile memory in the device):
16-247
•
In PROGRAM mode, navigate to:
Level Meter > Communication (5.) > Device Address
(5.2.)
to view the device address.
•
Press
Mode
to return to Measurement mode.
6.3
Application examples
Note
In the applications illustrated below, values are for example purposes only.
Содержание SITRANS L
Страница 258: ...LCD menu structure SITRANS LR250 with FOUNDATION FIELDBUS 258 Operating Instructions 09 2021 A5E32221411 AH ...
Страница 259: ...LCD menu structure SITRANS LR250 with FOUNDATION FIELDBUS Operating Instructions 09 2021 A5E32221411 AH 259 ...
Страница 260: ...LCD menu structure SITRANS LR250 with FOUNDATION FIELDBUS 260 Operating Instructions 09 2021 A5E32221411 AH ...