Modbus holding registers
A.3 Modbus holding registers tables
FC410 with Modbus
138
Operating Instructions, 02/2016, A5E33120874-AC
A.3.10
Quality codes for process values
Table A- 20 Quality code for process values
Modbus
address
Data type /
Size [bytes]
Parameter
Description
Default value [unit]
Value range
Access
level
3014
Unsigned /
4
Alarm Group 2
Quality code of a
measured value
Process values for
quality codes
Media temperature
Density
Volume flow
Mass flow
Quality code for each
process value consist
of 2 bits:
Bit 24/25:
Media temperature
Bit 26/27:
Density
Bit 28/29:
Volume flow
Bit 30/31:
Mass flow
11 Good
01 Reserved
10 Simulation
00 Bad
Read Only