Technical data
11.9 Power supply
FC410 with Modbus
100
Operating Instructions, 02/2016, A5E33120874-AC
Description
Specification
Measuring tube surface roughness
•
Standard: 1.6 µm
•
Hygienic: 0.8 μm
Self-draining design
Yes, when mounted vertically
*: Pressure ratings depend on sensor material
**: Pressure ratings depend on process connection dimension
11.9
Power supply
Table 11- 26 Power supply
Description
Specification
Supply Voltage [V]
24 VDC +/- 20% for non-hazardous duty
24 VDC +0/-20% for flameproof installation
Reverse polarity protection
Yes
Power consumption
1.1 W
11.10
Basic electrical requirement for master system
Table 11- 27 Power unit
Description
Specification
Protection class
Class I (EN60950)
Galvanic isolation primary/secondary
SELV acc. to EN 60950 and EN 50178 or equiva-
lent
Master/Receiver
Isolation 500 VAC
Содержание SITRANS FC410
Страница 2: ......
Страница 70: ...Commissioning 6 14 Process variables FC410 with Modbus 68 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 92: ...Troubleshooting FAQs 10 2 Troubleshooting FC410 with Modbus 90 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 142: ...Sensor dimension dependent default settings FC410 with Modbus 140 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 146: ...Zero point adjustment FC410 with Modbus 144 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 150: ...CRC calculation FC410 with Modbus 148 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 152: ...Exception codes E 1 Exception handling FC410 with Modbus 150 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 154: ...Float definition F 1 Float definition FC410 with Modbus 152 Operating Instructions 02 2016 A5E33120874 AC ...