Zero point adjustment
D
In the following the automatic zero point adjustment function is described in detail.
Note
Preconditions
Before a zero point adjustment is initiated, the pipe must be flushed, filled and at an absolute
flowrate of zero preferably also at operating pressure and temperature. Refer to "Zero point
adjustment" via HMI (Page 86) or SIMATIC PDM (Page 102) for instructions.
Note
Change of parameters during zero point adjustment
Do not change any other parameter during the zero point adjustment procedure.
Automatic zero point adjustment
SITRANS FC330 measures and calculates the correct zero point automatically.
The automatic zero point adjustment of the flowmeter is set by the following parameters:
● Zero Point Adjustment Period
● Start Zero Point Adjustment
When zero adjust is initiated by selecting "Start Zero Point Adjustment", the mass flow values
are acquired and totalized for the configured period (Zero Point Adjustment Period). The default
zero point adjustment period (30 s.) is normally sufficient for a stable zero point measurement.
Note
Extremely low flow quantity
If the flow quantity is extremely small, extremely precise measurement is necessary. In this
case, a long zero point adjustment period can be selected for improved zero point adjustment.
Zero point calculation
During zero point adjustment, an average value is automatically calculated using the following
formula:
Zero Point Offset Value
Average of N flow values
N
x
x
N
i
i
∑
=
≡
1
x
i
is an instantaneous flow value sampled in
the time domain
N = Number of samples during zero point ad‐
justment
SITRANS FC330
Operating Instructions, 01/2019, A5E44030648-AB
271
Содержание SITRANS FC330
Страница 14: ...Introduction 1 8 Notes on warranty SITRANS FC330 14 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 56: ...Installing mounting 4 3 Disassembly SITRANS FC330 56 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 126: ...Operating 7 2 Remote operation SITRANS FC330 126 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 158: ...Parameter assignment 8 2 Functions SITRANS FC330 158 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 216: ...Technical data 11 13 Pressure temperature ratings SITRANS FC330 216 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 224: ...Dimension drawings 12 7 Mounting bracket dimensions SITRANS FC330 224 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 236: ...HART communication B 1 Mode of operation HART function SITRANS FC330 236 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 274: ...Zero point adjustment SITRANS FC330 274 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 280: ...Index SITRANS FC330 280 Operating Instructions 01 2019 A5E44030648 AB ...