Manual zero point adjustment
In case an automatic zero point adjustment cannot be performed, it is possible to do a manual
zero point adjustment by entering the zero point offset value.
1. Select "Manual" in "Zero point adjustment" (menu item 2.1.3).
2. Enter the desired value in "Offset" (menu item 2.1.3.7).
8.2.3
Low flow cut-off
In certain applications, as for instance dosing applications, 0% flow signals below a certain
flowrate are desired. In these applications, the flow signal can be forced to zero, when the flow
is lower than a predefined flow value (Low Flow Cut-Off).
The device provides two parameters for setting the low flow cut-off:
● Low Mass Flow Cut-Off
● Low Volume Flow Cut-Off
The low flow cut-off parameters influence all outputs of the device, for example Local User
Interface, Channel 1 to 4, and bus communication outputs.
Depending on the process values selection of the output either Low Mass Flow Cut-Off or Low
Volume Flow Cut-Off will influence the output.
8.2.4
Empty tube monitoring
The empty tube monitoring function uses the process density for detecting an empty tube. Use
of this function is recommended for all standard applications.
Note
Gas applications
Deactivate the empty tube monitoring function.
Empty tube monitoring parameters
Two parameters for setting the empty tube monitoring function are available:
● Empty Tube Detection (Modbus address 2129)
● Empty Tube Limit (Modbus address 2127)
The empty tube monitoring is activated via the Empty Tube Detection parameter. When the
empty tube monitoring function is on, the mass flow / volume flow value is forced to zero if the
tube is empty.
Parameter assignment
8.2 Functions
SITRANS FC330
132
Operating Instructions, 01/2019, A5E44030648-AB
Содержание SITRANS FC330
Страница 14: ...Introduction 1 8 Notes on warranty SITRANS FC330 14 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 56: ...Installing mounting 4 3 Disassembly SITRANS FC330 56 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 126: ...Operating 7 2 Remote operation SITRANS FC330 126 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 158: ...Parameter assignment 8 2 Functions SITRANS FC330 158 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 216: ...Technical data 11 13 Pressure temperature ratings SITRANS FC330 216 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 224: ...Dimension drawings 12 7 Mounting bracket dimensions SITRANS FC330 224 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 236: ...HART communication B 1 Mode of operation HART function SITRANS FC330 236 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 274: ...Zero point adjustment SITRANS FC330 274 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 280: ...Index SITRANS FC330 280 Operating Instructions 01 2019 A5E44030648 AB ...