Modbus communication
A.9 Modbus holding registers tables
SITRANS F Ultrasonic flowmeters FST030 (Modbus)
Operating Instructions, 08/2017, A5E37367739-AC
181
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[units]
(units regis-
ter)
Value range /
Setting options
Access
level
7403
Float / 4
Upper warning
limit
Exceeding this limit
causes a warning.
max. mass
flow
[kg/s]
(7400)
-max. mass flow -
+max. mass flow
Read /
write
7405
Float / 4
Lower warning
limit
Falling below this limit
causes a warning.
-max. mass
flow
[kg/s]
(7400)
-max. mass flow -
+max. mass flow
Read /
write
7407
Float / 4
Lower alarm
limit
Falling below this limit
causes an alarm.
-max. mass
flow
[kg/s]
(7400)
-max. mass flow -
+max. mass flow
Read /
write
7412
Float / 4
Hysteresis
Hysteresis that affects
the alarm and warning
limits.
0
[kg/s]
(7400)
0 - max. mass flow
Read /
write
Standard volume flow settings
Table A- 31 Standard volume flow settings (only hydrocarbon applications)
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[units]
(units regis-
ter)
Value range /
Setting options
Access
level
1723
Float / 4
Standard condi-
tion tempera-
ture
Specifies temperature
for standard conditions.
The value is used to
calculate standard
process values.
15.0
[°C]
(7700)
0.0 - 100.0
Read /
write
1725
Float / 4
Standard condi-
tion pressure
Specifies pressure for
standard conditions.
The value is used to
calculate standard
process values.
101325
[Pa]
(8015)
100000.0 - 200000.0
Read /
write
1731
Float / 4
Standard condi-
tion air density
Specifies air density for
standard conditions.
The value is used to
calculate standard
process values.
1000.0
[kg/m
3
]
(7600)
0.080 - 10000.0
Read /
write
7531
Float / 4
Low flow cut-off Standard volume flow
limit for low flow cut-off.
Below limit standard
volume flow output is
forced to zero.
0.0
[Sm
3
/h]
(7964)
0 - max. standard vol-
ume flow
Read /
write