MAG 6000 I / MAG 6000 I Ex de transmitter
Operating Instructions, 04/2016, A5E02083319-AG
89
Factory settings
B
B.1
Transmitter factory settings
The transmitter is delivered with the following factory settings:
Menu item
Parameter
Factory
settings
Options
More info
Password
Password
1000
1000 ... 9999
(Page 42)
Basic settings
Flow direction
Positive
Positive, negative
(Page 42)
Q
max
Sensor size
dependent
Sensor size dependent
•
volume unit
Sensor size
dependent
m
3
, ml, l , hl, kl, Ml, ft
3
, in
3
, US G, US kG, US MG,
US BBL, UK G, UK MG, ? (customer unit)
•
time unit
Sensor size
dependent
Sec., min., hour, day, ? (customer unit)
Totalizer 1
Forward
Forward, reverse, net
•
Totalizer 1 unit
Sensor size
dependent
m
3
, ml, l , hl, kl, Ml, ft
3
, in
3
, US G, US kG, US MG,
US BBL, UK G, UK MG, ? (customer unit)
Totalizer 2
Reverse
Forward, reverse, net
•
Totalizer 2 unit
Sensor size
dependent
m
3
, ml, l , hl, kl, Ml, ft
3
, in
3
, US G, US kG, US MG,
US BBL, UK G, UK MG, ? (customer unit)
Low flow cut-off
1.5%
0 ... 9.9%
Empty pipe
Off
On, Off
Velocity unit
m/s
m, mm, cm, dm, ft , in
per s, min, h, d, ? (customer unit)
Error level
Warning
Fatal, permanent, warning
Output
Current output
On
On/off, Unidirectional/bidirectional, 0...20 mA/
4...20 mA/4...20 mA + Alarm
tings (Pa-
ge 47)
•
Alarm level
No
Yes/No
•
Alarm diff.
Low
High/Low
•
Time constant
5 s
0.1...30 s
Digital output
Pulse
Error, direction/limit, batch, frequency, pulse, error
number, off
(Page 77)
Relay output
Error level
Error, direction/limit, cleaning, error number, off
(Page 77)
Direction/limit
switch
Off
1 setpoint, 2 setpoints
(Page 78)
•
Setpoints
0%
-100 ... +100%
•
Hysteresis
0%
0.0 ... 100%
Содержание SITRANS F MAG 6000 EX DE