Type I intermittent checking facilities are automatic checks performed at certain time intervals
or per fixed number of measurements. They include:
● Checksum calculation (10 min. interval on totalizer checksum)
● Insulation test (minimum 24 hour interval)
● Battery capacity check (4 hours interval)
Type N non-automatic checking facilities are checks that are not performed automatically
including all other diagnosis functions in the MAG 8000 and the MAG 8000 CT.
9.3
Flow simulation
MAG 8000 and MAG 8000 CT has a built-in flow simulator (parameters 551 and 552) to verify
and adjust pulse output to any connected device or system.
WARNING
Totalized values are changed during simulation and actual flow is NOT measured.
Simulation will stop automatically in 4 hours.
For MAG 8000 CT flow simulation is only possible when verification sealings are broken.
Troubleshooting/FAQs
9.3 Flow simulation
SITRANS MAG 8000
Operating Instructions, 11/2018, A5E03071515-AE
89
Содержание SITRANS F M MAG 8000
Страница 10: ...Introduction 1 3 Further Information SITRANS MAG 8000 10 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 16: ...Description 3 4 Benefits SITRANS MAG 8000 16 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 56: ...Commissioning 6 6 Data protection SITRANS MAG 8000 56 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 74: ...Operation 7 6 Battery powered operation SITRANS MAG 8000 74 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 90: ...Troubleshooting FAQs 9 3 Flow simulation SITRANS MAG 8000 90 Operating Instructions 11 2018 A5E03071515 AE ...
Страница 164: ...Index SITRANS MAG 8000 164 Operating Instructions 11 2018 A5E03071515 AE ...