Menu
Parameter
Display example
Comments
Menu 4
Alarm codes
Each code refers to a specific
alarm.
Menu 5
Display test
Check of all segments. Dis‐
play toggles between all seg‐
ments on/off.
6.3
Start-up routine
1. Power-up device.
2. Reset battery status indicator as described below.
Resetting the battery counter
When a new battery is installed and the plug is connected, the transmitter start-up routine
begins. The display shows the active software version, e.g. 2.03.
After ten seconds the message "reset.bat" will appear.
1. Press push button within six seconds to reset the internal battery counter. The message
"accept" will appear.
2. Press push button again within six seconds in order to reset the internal battery counter.
The battery indicator now shows full. If the push button is not pressed again, the battery
indicator will show "Low".
/
Figure 6-2
Reset internal battery counter
* Firmware version, here version 2.03
Commissioning
6.3 Start-up routine
FUS380/FUE380
Operating Instructions, 09/2016, A5E00730100-AB
47
Содержание SITRANS F FUE380
Страница 6: ...Table of contents FUS380 FUE380 6 Operating Instructions 09 2016 A5E00730100 AB ...
Страница 20: ...Description 3 5 Principle of operation FUS380 FUE380 20 Operating Instructions 09 2016 A5E00730100 AB ...
Страница 62: ...Functions 7 4 Hardware key FUS380 FUE380 62 Operating Instructions 09 2016 A5E00730100 AB ...
Страница 78: ...Troubleshooting FAQs 9 2 Diagnosing with PDM FUS380 FUE380 78 Operating Instructions 09 2016 A5E00730100 AB ...
Страница 100: ...Settings B 3 Ordering of spare parts FUS380 FUE380 100 Operating Instructions 09 2016 A5E00730100 AB ...