Connecting
5.5 Finishing the transmitter connection
SITRANS F Ultrasonic flowmeters FST030 (HART)
Operating Instructions, 08/2017, A5E35971532-AC
49
5.
Connect ground to terminal and power to terminals L/+ and N/- in the manner shown
below at right using a screwdriver.
①
L/
+
②
N/
-
③
Protective Earth (PE)
AC connection
DC connection
Power: 85 to 264 V AC, 47 to 63 Hz
Power: 19.2 to 28.8 V DC
6.
Mount power supply protection cover and fasten protection cover screw.
7.
Tighten cable gland.
5.5
Finishing the transmitter connection
Connection check-up
1.
Check individual wire installation by tugging firmly.
2.
Firmly tighten cable glands and insert blanking plugs in unused cable entries.
3.
Close lid.
4.
Tighten the four spring screws.
5.
Ensure that moisture does not penetrate to inside of electronics enclosure.
Your device is now ready for commissioning.