![Siemens SITRANS 7MP1110-0AC Скачать руководство пользователя страница 67](http://html.mh-extra.com/html/siemens/sitrans-7mp1110-0ac/sitrans-7mp1110-0ac_operating-instructions-manual_384651067.webp)
Operation
5.12 Increase service life
SITRANS TF280 WirelessHART
Operating Instructions, 10/2010, A5E02619566-02
67
5.12
Increase service life
Requirement
● You have executed the quick start function: Quick start (Page 62)
● You are acquainted with the principles of operation: Use of local user interface (Page 38)
Procedure
1.
Switch off HART maintenance port.
Parameter:
– "1.5 HART modem" (Page 45)or
– "4.2.1 HART modem" (Page 56)
Switching-off the HART maintenance port reduces the energy consumption and thus
reduces the battery service life.
2.
Set "Burst display" function.
Parameter: "2.1.2 Burst display" (Page 46)
The battery service life is highest if both checkboxes are deactivated.