Replacing boards / replacing components
6 - 17
Siemens AG
SPR2-230.061.01
Page 17 of 48
SIREMOBIL Iso-C
Medical Solutions
Rev. 13
12.04
CS PS 24
System Manual
Replacing the Compact optics
6
Removal
Camera optics
6
•
Move the I.I. into the service position. Refer to chapter 1, I.I. service position.
•
Rotate the camera into the 0° position. Refer to Fig. 6.
Fig. 6
Electrical voltage!
See Chapter 1, Safety Information.
Switch the SIREMOBIL Compact off and wait approximately
3 minutes for the voltage in the I.I. mini voltage supply to drop.
Risk of injuries on mechanical parts!
If not observed, slight to medium injuries, especially to the hands,
can occur.
Observe the notes concerning the C-arm carbon fiber structure.
Risk of injury!
Noncompliance can lead to light to medium injuries and/or
property damage.
The C-arm covers contain counterbalance weights and weigh
approx. 6 kg.
Hold the covers firmly when installing or removing them.
Before the C-arm cover will be loosened, the C-arm must be
moved to its middle orbital position.
If the C-arm covers are loosened, the C-arm may not be moved to
its orbital end positions (I.I. or SIREPHOS at the orbital end posi-
tion).
0° Position
WARNING
WARNING
CAUTION