3. Part of activities independent of serial number
•
Connect the S-Video Out plug with the terminator (see Fig. 73).
•
Do not use the terminator for 3D software, version VA00A. Instead, tape the terminator to
the PC housing.
•
Use the terminator for the following 3D software .
•
Secure wth cable ties.
Cable no. 14, item number n. a.
6
is included 2x in system cabling 71 39 855
1. Cable
•
Connect one plug (see Fig. 59) at input 2G of video switcher 2 (see Fig. 69).
•
Pull the other plug down through the left side panel of the monitor trolley.
•
Connect this plug with BNC coupling/ Cinch (see Fig. 73).
2. Cable
•
Disconnect the existing Out3 plug at the Memoskop. Feed it through the side opening,
pull it upward and insert it at Out G on video switcher 2 (see Fig. 69).
•
Connect one plug of the second cable no. 14 to input 1G of video switcher 2
(see Fig. 69) .
•
Route the other plug down to the Memoskop and insert it into Out3.
•
Slide the excess cable into the side panel.
Fig. 75 Cable no.14
Markierung für Tastaturschublade