Replacing boards / replacing components
Power board D2
6
•
Disconnect boards D1 and D2 completely.
•
Disconnect the cables between boards D1 and D2.
•
Remove board D1 and D2.
•
Apply heat conducting paste to the heat sink for the power semiconductor on the new D2
board.
•
Reinstall board D2 and D1 and reattach all connections.
•
Ensure that the shielding and ground connections are positioned correctly.
•
Perform the kV offset adjustment and the generator adjustment (filament learning).
•
Test the FL / PFL / DR and direct exposure functions (if there is a cassette holder).
Interface board D3
6
•
Replace interface board D3.
•
Check the +26.75 V voltage for the I.I. mini voltage supply. Adjust it if necessary.
Refer to Chapter 3 of these instructions and check the operating voltages.
•
Check the radiation release / format switch over / vertical column travel functions.
D40 board for downward travel of the lifting column, Compact L
only
6
•
Replace the D40 board.
•
Using the operating instructions for the basic system, check both the movements of the
C-arm and the signal messages.
Electrical voltage!
See Chapter 1, Safety Information.
Make sure that LED V400 is off prior to performing any work on
boards D1 and D2. This should occur approximately 3 minutes
after switching the SIREMOBIL Compact OFF.
Electrical voltage!
See Chapter 1, Safety Information.
Switch the system power supply off and disconnect the power
plug.
Electrical voltage!
See Chapter 1, Safety Information.
Switch the system power supply off and disconnect the power
plug.
WARNING
WARNING
WARNING