SIREMOBIL Compact
SPR2-130.061.02
Page 26 of 30
Siemens AG
System Manual
Rev. 11
09.04
CS PS 24
Medical Solutions
5 - 26
Adjustments / Programs
Overview of SIREMOBIL Compact parameters
5
Invoke the Moreinfo.hlp help file on the service PC for a short description of the parame-
ters available on the SIREMOBIL Compact.
Overview of MEMOSKOP C / C-E / C-SUB parameters
5
User Setup
•
The User Setup is invoked with the CTRL + U key combination.
•
The available functions are described in the SIREMOBIL Compact Operating Instruc-
tions.
Technical Setup
•
The technical setup is invoked with the CTRL + T key combination.
•
Enter the password and confirm with [RETURN]
The following functions can be invoked depending on the type of MEMOSKOP:
If parameters have been changed and saved in the
"Adjustments, Parameter" menu with Save to file, you must exit
the service program with "Logoff" and "Quit" to save the modified
parameters to the diskette.
Function:
Comments:
Test pattern generation
Not for service, invoke the necessary test images on the
service PC.
Video frequency
The currently programmed frame rate is shown in the
parentheses to the right. After downloading the
MEMOSKOP or after replacement, the MEMOSKOP will
start up at the default video frequency (50 Hz).
If the monitors are configured for a different video fre-
quency, you must perform "blind" configuration for the cor-
rect video frequency.
Refer to Configuring video frequency.
Number of monitors
Allows you to program the number of monitors used
(1 or 2).
Hardcopy setup
You can configure the connection for hardcopy cameras.
Check whether 5000 is programmed for "4 Max Hardcopy
time" and if not, program this value.
Edge 1 coef set
Allows configuration of edge enhancement. Check
whether 3 is programmed and if not, program this value.
Edge 2 coef set
Allows Configuration of edge enhancement.Check
whether 8 is programmed and if not, program this value.
Motion detection setup
Configures the motion detector.
Check if at MD 1: Threshold 100 and at MD 2:
Threshold 120 is programmed, if not, program these val-
ues.
Circle mask setup
Allows you to configure the diameter of the circle mask.
NOTE