Adjustments / Programs
5 - 19
Siemens AG
SPR2-130.061.02
Page 19 of 30
SIREMOBIL Compact
Medical Solutions
Rev. 11
09.04
CS PS 24
System Manual
Correction of the 0° position of the slot diaphragm display
5
•
Select the "Adjustments," "Parameter" menu at the service PC.
•
Select "Screen display Slot diaphragm" in the Combo box.
•
Click "Get from Unit".
•
Select "Rotation 0 degree position".
•
The parameter in the "Value Actual" field can be changed according to the minimum and
maximum values.
•
Then click "Set Value".
•
Click "Put to Unit".
•
After the SIREMOBIL Compact has booted, compare the 0° position of the slot dia-
phragm blades to the display on the monitor. If necessary, repeat the programming.
•
Click "Save to file" .
•
Leave the service program with "Logoff" and "Quit" to save the parameters as a new file
to disk.
Correct only if the slot diaphragms are shown at an angle of rota-
tion different from that displayed on the monitor. A correction of
+ 10° is possible. Camera rotation and slot diaphragm have to be
set correctly.
The service PC is connected to the serial interface of the
SIREMOBIL Compact.
The service program is started.
NOTE