
3 - 1
Siemens AG
SPR2-130.061.02
Page 1 of 2
SIREMOBIL Compact
Medical Solutions
Rev. 11
09.04
CS PS 24
System Manual
Checking the operating voltages
3
Low voltages
3
Check the operating voltages for the SIREMOBIL Compact against the values in Table 1.
I.I. voltages
3
The voltages E1 / E2 / E3 and A may be taken from the I.I. test protocol and checked or
adjusted against the control test points listed in Table 2.
from test point
to test point
Voltage
Tolerance
Potentiometer
M14.4 (0V)
M14.5/6 (+5 V)
+ 5 V
+ 0.2 V
M14. +5 V/Adj.
M14.3 (0V)
M14.2 (+15 V)
+ 15 V
+ 0.1 V
M14. +15 V/Adj.
M14.3 (0V)
M14.1 (-15 V)
- 15 V
+ 0.5 V
M14. -15 V/Adj.
D3.X8.1 (0V)
D3.X8.3 (+26,75 V)
+26.75 V
+ 0.25 V
D3.R3
M13.S- (0V)
M13.S+ (+13 V)
+13 V
+ 0.2 V
M13.TR1
D3.X1.7 (0V)
D3.X1.9 (+24 V)
+ 26.5 V to
+ 29.5 V
n.a.
not adjustable
Tab. 1
Electrical voltage!
See Chapter 1, Safety Information.
Do not touch any part of the I.I. mini voltage supply!
Voltage
Test point
Ground
point
(0V)
Potentiome-
ter for full
format
Potentio-
meter
for zoom
format
Voltage
divider
ratio
E1
UE1
P10
P11
1:1
E2
UE2
P6
P7
1:1
E3
UI 15
P2
P3
1:10000
30 kV, Anode
UI 30
P1
P1
1:10000
Tab. 2
WARNING