1 - 1
Siemens AG
SPR2-130.814.01
Page 1 of 2
SIREMOBIL C
Medical Solutions
Rev. 01
03.01
TD PS 24
System Manual
General
1
Safety notes
1
The product-specific safety notes specified in the system manual and the safety notes
specified in ARTD Part 2 must be observed.
After all work has been completed and after all covers have been fitted, the protective
conductor test must be carried out in accordance with ARTD-002.731.17. The protective
conductor resistance must not exceed 0.2 Ohms.
Required documents
1
Required measuring equipment
1
Required tools
1
•
Operating Instructions Codonics Printer / Framegrabber
•
System Folder for Basic Unit
•
Densitometer X-RITE 331
97 02 416 Y1996
•
Protective conductor meter
or
44 15 899 RV090
•
Bender safety tester
99 00 440 RE999
•
Tool kit
•
1 set of Allen keys
•
1 set of torque screwdrivers