Commissioning
3 - 9
Siemens AG
SPR2-130.814.01
Page 9 of 10
SIREMOBIL Compact
Medical Solutions
Rev. 01 03.01
TD PS 24
System Manual
Final work
3
•
File the installation and configuration instructions in the system manual.
•
Refit all covers.
•
Carry out the protective conductor test in accordance with ARTD-002.731.17. The
protective conductor resistance must not exceed 0.2 Ohms.
•
Hand over the operating instructions to the customer.