SIREMOBIL Compact
SPR2-130.814.01
Page 4 of 10
Siemens AG
System Manual
Rev. 01
03.01
TD PS 24
Medical Solutions
3 - 4
Commissioning
MEMOSKOP Video singals at hardcopy output
3
Default programming: Without pressing the Hardcopy key at the unit, the image from the
live monitor is transferred to the Memoskop video output OUT3. With pressing the Hard-
copy key at the unit, for a short time, the image from the live monitor is transferred to the
Memoskop video output OUT3.
Changes can be performed according to the description in the help file of the Memoskop.
Read out of the actual video frequency
3
•
Read out the programmed video frequency on the MEMOSKOP and note down.
- Call up the Technical Setup if a MEMOSKOP keyboard is available. For this,
simultaneously press the CTRL + T keys and enter the password.
Read out the video frequency in the Video Frequency line and note down. Do not close
the Technical Setup yet.
- If no keyboard is available, connect the Service PC to the system and start the
MEMOSKOP service program.
- In the main menu, press the "Get from Memoskop" button. Call up the "Others" menu,
read out the video frequency and note down. Do not exit the Technical Setup yet.
Memoskop SMPTE test image selection
3
•
Select LUT 1 for both monitors on the MEMOSKOP keyboard or on the basic unit.
•
Call up the SPMPTE test image on the MEMOSKOP.
- If a MEMOSKOP keyboard is available, call up the SPMPTE test image in the User
Setup and display on both monitors.
- If no keyboard is available, select the menu "Test" and select "SMPTE" in the service
program.
- Then click the "Put to Memoskop" button in the main menu of the service program. The
SMPTE test image is displayed on both monitors.
•
If a MEMOSKOP keyboard is available, switch on the text overlay.
The default programmings lead to the following effect:
With the release of the film print, using the hardcopy keys of the
unit, the image of the reference monitor will be printed.
With the release of the film print, using the hand-held control of
the framegrabber, the image of the live monitor will be printed.
NOTICE