Important functions (selection)
7.5 Parameter assignments
SIPROCESS UV600
140
Operating Instructions, 10/2013, A5E03529375-02
Selecting the signal type (category) for the limit message
Use the
,
and
keys to select the fourth parameter "Type".
Set the signal type in the follow-up menu using the
,
and
keys. You can
select the following:
Status (flag) Description Possible cause
F
Fault
Internal fault/malfunction (unsafe operating state)
M
Maintenance Maintenance request/maintenance status
C
Check
An internal function is currently running, e.g. a calibration. Measuring function is interrupted
U
Uncertain
Unsafe operating state or measured value
E
Extended
Internal message
Selection of digital output for limit message
Use the
,
and
keys to select the fifth parameter "Status".
A menu is displayed with the available digital outputs. Use the
,
and
keys to
select the desired digital output.
A further menu is displayed in which you can select the electronic switching logic (direct or
inverted).
7.5.7
Pump settings
Switching the pump on and off
This function can be used to manually switch the sample gas pump (option) on and off. This
function controls the integral sample gas pump and the associated tag BVO1. The pump
cannot be switched on if it is switched off for other reasons.
The pump remains switched off automatically
●
if the temperature of the measuring system is too low following switching-on of the
analyzer
●
if the temperature of the measuring system is too low following switching-on of the
analyzer
●
if the condensation sensor (option) has been triggered
●
during the flow of calibration gas if this is set in this manner
●
if a digital input is configured with the "Pump off" function and activated
Menu call:
Main menu → Parameters → Pump → Pump off