Important functions (selection)
7.3 Calibration and validation functions
SIPROCESS UV600
118
Operating Instructions, 10/2013, A5E03529375-02
The following menu is displayed (example):
Function 1 (relative) is not provided for the validation measurement. The list therefore
commences with 2. Meaning of the various parameters:
Display on screen
Description
Date
Current date in representation YY-MM-DD
Time
Current time in 24-hour representation
Setpoint
Setpoint of the used calibration gas
Actual value
Mean value of the measured actual values
7.3.7
Group functions
This function can be used to manually start preconfigured, automatic calibration procedures.
Usually a number of calibration procedures which match the individual analyzer are specified
by the manufacturer.
Only those calibration procedures are available which can be carried out in the current
operating mode. If a status message exists whose cause would make a calibration
procedure unreliable or impossible, the associated calibration procedures cannot be
selected. Calibration procedures which are currently "active" cannot be started.
Menu call:
Main menu → Calibrations → Group functions