3. Wait for transition to warming-up phase:
When the splash screen disappears, the device is in the warming-up phase. The current
temperature status is displayed as a bar graph in the warming-up window.
4. Call the measured-value display during the warming-up phase
– Close the warming-up window by pressing <ENTER>.
The display shows the measured-value display in read mode.
For the duration of the warming-up phase, the device is in "Function check" mode. The
following symbol is displayed in the status bar:
Setting up
Note
Operator control options
When commissioning for the first time, two methods can be used. You can commission the
device for the first time with the procedure described below or you can use a simplified
procedure.
Information on simplified procedure → Table A-3 References 1 - Operating Manuals LUI
(Page 99). Make the necessary settings in the appropriate submenus.
If you do not use the quick start procedure, commission the device with the steps described
below:
1. Call the main menu:
– Switch the measured-value display to selection mode.
To do so, press <ENTER>.
– Press <Enter> again to open the main menu.
2. Identify device: [3.09] Maintenance & Diagnostics > Identification
3. Define PIN for the user levels: [5.1] Security > Access management
In the factory settings, the PIN 1111 is the default for the "Standard" user level and PIN
2222 for the "Expert" user level.
4. Disable automatic log-off temporarily:
The factory setting is a log-off time of 5 min. When this time elapses, changes that were
not stored retentively are lost. We recommend that you temporarily deactivate the automatic
log-off during commissioning.
Commissioning
8.4 Commissioning the SIPROCESS GA700
Field device
76
Operating Instructions, 05/2015, A5E35640463-02
Содержание SIPROCESS GA700
Страница 6: ...Inhaltsverzeichnis Feldgerät 6 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 22: ...Allgemeine Sicherheitshinweise 2 9 Automatisches Reset Feldgerät 22 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 64: ...Anschließen 6 5 Feldbediengerät Feldgerät 64 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 66: ...Bedienen 7 1 Bedienen Feldgerät 66 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 94: ...Technische Daten 10 3 Technische Daten OXYMAT 7 Feldgerät 94 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 112: ...Index Feldgerät 112 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 118: ...Table of contents Field device 6 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 134: ...General safety instructions 2 9 Automatic reset Field device 22 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 158: ...Installing mounting 5 3 Field module Field device 46 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 176: ...Wiring 6 5 Field control unit Field device 64 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 178: ...Operation 7 1 Operation Field device 66 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 190: ...Commissioning 8 4 Commissioning the SIPROCESS GA700 Field device 78 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 222: ...Appendix A 6 Returned delivery Field device 110 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 224: ...Index Field device 112 Operating Instructions 05 2015 A5E35640463 02 ...