8.3.4
Reference gas from the gas cylinder
For a sufficient measuring ability, the reference gas pressure must be at least 2 000 hPa +/-
150 hPa above the sample gas pressure.
1. Use a reference gas line with the following characteristics at measuring spans of < 2%:
– Small cross-section, e.g., pipe diameter 1/16"
– Made of metal, e.g., stainless steel
2. Flush the reference gas line thoroughly with reference gas before commissioning.
3. Check the reference gas line for leaks to avoid unnecessary losses:
– Close the valve at the gas cylinder.
– If the pressure display on the pressure reducer does not drop by more than 1 000 hPa/
min, the gas connection is sufficiently tight.
In the case of two-stage pressure reducers, the inlet pressure must not drop by more
than 1 000 hPa/min.
8.3.5
Checking the reference gas flow
1. Close the sample gas inlet.
2. Lay a hose with an internal diameter of 4 mm from the sample gas outlet into a beaker filled
with water.
3. Apply reference gas to the analyzer module.
The reference gas must exit slowly with 1 to 2 bubbles per second through the water filling (at
versions with flow-type compensation circuit, 2 to 4 bubbles per second). The reference gas
flow rate is then sufficient.
8.3.6
Sample gas inlet conditions
Prevent condensation of the sample gas in the gas path by using suitable measures.
Commissioning
8.3 OXYMAT 7
Field device
Operating Instructions, 05/2015, A5E35640463-02
73
Содержание SIPROCESS GA700
Страница 6: ...Inhaltsverzeichnis Feldgerät 6 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 22: ...Allgemeine Sicherheitshinweise 2 9 Automatisches Reset Feldgerät 22 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 64: ...Anschließen 6 5 Feldbediengerät Feldgerät 64 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 66: ...Bedienen 7 1 Bedienen Feldgerät 66 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 94: ...Technische Daten 10 3 Technische Daten OXYMAT 7 Feldgerät 94 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 112: ...Index Feldgerät 112 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 118: ...Table of contents Field device 6 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 134: ...General safety instructions 2 9 Automatic reset Field device 22 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 158: ...Installing mounting 5 3 Field module Field device 46 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 176: ...Wiring 6 5 Field control unit Field device 64 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 178: ...Operation 7 1 Operation Field device 66 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 190: ...Commissioning 8 4 Commissioning the SIPROCESS GA700 Field device 78 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 222: ...Appendix A 6 Returned delivery Field device 110 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 224: ...Index Field device 112 Operating Instructions 05 2015 A5E35640463 02 ...