5.7 Input / Output Modules
99
Power Meter SICAM P 7KG7750/55, Manual
E50417-B1076-C340-A6, Edition 09.2013
5.7.6
Binary Inputs (optional)
The binary inputs (optional) can be used for static messages and impulse inputs.
Proceed as follows:
1. Activate/Deactivate one or both binary inputs by ticking the corresponding check box under
1.
and/or
2.
2. For each input used, set the
Designation
together with the
Dimension
.
3. If you want to use the input as an external counter, you have to define the
Factor
e.g. as
Energy increase per pulse (refer to chapter 5.7.2).
4. If you want to use the binary input to display binary information (0/1), 0 must be entered as
a
factor
.
5. Choose the binary input to be used for time
Synchronization
via minute impulses. The
device will display the designation, not the number of the
Binary input
.
6. Analog to measured values, binary inputs may be displayed on measurement screens (see
Note
The designation of the binary input is displayed instead of the binary input number!