3
Servo Ventilator 300/300A
Disassembling and assembling
E382 E380E 061 01 03 01
Siemens-Elema AB
93
Separating the pneumatic
section from the power
section
• Place the patient unit on the edge of the
workbench so the screws are accessible
from below.
• Loosen the four screws (1).
• Lift off the pneumatic section from the
power section. Place the pneumatic
section on a support beside the power
section.
• Disconnect the battery connector (2).
To reduce the risk of damages, caused
by short-circuits in the power section,
always have the batteries disconnected
when the power section is open.
• Instead of cutting the cable ties, remove
the screw (3) holding the battery
bracket (4).
• Disconnect the connectors (5, 6 and 7).
The pneumatic section is now separated
from the power section.
1
5
6
7
3
4
300-A35X
2