DU-DR-0016-SC9000 Rev A 6/4/02 GCX Corp. Page 1 of 4
GCX Mounting Assembly Operation/Installation Manual
Dräger Model Cato
®
with Siemens SC9000 Patient Monitor
Patient Monitor – Left Side on Pivot Arm / Power Supply on Right Side
The purpose of these instructions is to:
1. Describe attachment of GCX mounting hardware to the Cato anesthesia machine.
2. Describe attachment of the Siemens monitor to the GCX mounting hardware.
WARNING
: USE OF MOUNTING HARDWARE AND MONITORING COMPONENTS OTHER THAN THOSE
DESCRIBED
IN
THIS
DOCUMENT MAY RESULT IN SERIOUS INJURY DUE TO TIPPING OF THE
ANESTHESIA
MACHINE.
THIS CATO MOUNTING CONFIGURATION REQUIRES THE 40KG COUNTERWEIGHT WITH
EXTENDED
WHEEL
ADAPTER
UPGRADE
SET FROM DRÄGER (PART NUMBER: 86 018 68)
Installer
: When installation is completed, provide these instructions to the end user for future reference.
INSTALLATION
Tools Required: Philips head screw driver, Hex key (incl.).
Mounting the Channel to Cato Edition
Parts Required: 4" Channel, M6 x 16 mm FHMS screws (2).
Brass
Plate
with
inserts
from
Dräger
(M29110)
1. To attach the Channel to the Cato left side accessory track,
locate the brass plate in the accessory track, align the
mounting holes in the Channel with the threaded holes on
the brass plate then run the M5 flat head screws (2) partly
in. (Note: Fixed Stop is at the bottom of the Channel)
2. Set the height of the Channel in the accessory track and
tighten the M6 screws to hold in place.
Prepare M Series Arm
1) Install the Tilt Limiting Stop screw.
M6 x 16 mm (2) FHMS
Accessory Track
Fixed Stop
5/16-18 x ½” SHCS (1)