Rapidlab 1200 Operator’s Guide: Maintenance
5-25
02087462 Rev. V
Cleaning the CO-ox Roller Cage
Use this procedure to clean the roller cages of the CO-ox pump.
You need the following materials for this task:
•
Lint-free tissue and swabs
•
10% bleach solution
•
Water
1. Select
Status > Maintenance
.
2. Lift the front cover.
Refer to
Preparing for Maintenance Procedures
3. Disconnect the reagent manifold tubing from the CO-ox pump tubing at the
fitting.
Figure 5-12 CO-ox Pump Tubing to Reagent Manifold
4. Grasp the lower tubing cuff and gently pull it away from the platen.
BIOHAZARD:
Wear personal protective equipment. Use universal precautions.
Refer to
Appendix A, Protecting Yourself from Biohazards
for recommended
precautions when working with biohazardous materials.
1
Fitting
2
Reagent manifold tubing
Содержание Rapidlab 1200
Страница 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Страница 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Страница 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Страница 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Страница 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Страница 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Страница 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Страница 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Страница 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Страница 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...