Rapidlab 1200 Operator’s Guide: Maintenance
5-7
02087462 Rev. V
6. Aspirate the conditioner:
a. Draw the conditioner solution into a syringe and insert the syringe into the
sample port.
b. Select
Continue
.
c. When prompted, remove the syringe and select
Continue
.
The conditioning cycle runs for 10 minutes.
7. If you have a Rapidlab 1260 or 1265 system, replace the test/blank sensors
with the biosensors.
8. Select
Return
.
9. Select
Analyze
.
10. If you have a Rapidlab 1260 or 1265 system, select
Yes
.
11. If you have a Rapidlab 1240 or 1245 system, perform a 2-point calibration.
Refer to
Performing Manual Calibrations
12. Analyze a minimum of 2 levels of quality control material to verify sensor
performance.
Checking the Level of Fill Solution
Use this procedure to ensure that the sensors contain the correct level of fill
solution.
The
p
O
2
and
p
CO
2
sensors do not require maintenance. Slight discoloration of the
fill solution in the
p
O
2
and
p
CO
2
sensors is normal.
The Glucose and Lactate biosensors do not require fill solution.
BIOHAZARD:
Wear personal protective equipment. Use universal precautions.
Refer to
Appendix A, Protecting Yourself from Biohazards
for recommended
precautions when working with biohazardous materials.
CAUTION:
Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the
measurement module frame to discharge static buildup.
1. Select
Status > Maintenance > Performing Sensor Maintenance
.
2. Select
Continue
.
3. Lift the door to the measurement module.
Refer to
Содержание Rapidlab 1200
Страница 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Страница 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Страница 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Страница 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Страница 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Страница 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Страница 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Страница 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Страница 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Страница 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...