5-4
Rapidlab 1200 Operator’s Guide: Maintenance
02087462 Rev. V
8. If the Glucose or Lactate results are below the values recommended on the High G/L
package insert, replace the affected biosensor.
Refer to
Performing Weekly Maintenance
Except as stated in the procedures, the weekly maintenance schedule is based on analyzing
30 samples per day, unless otherwise noted. If your laboratory analyzes more than 30
samples per day, perform this maintenance more frequently.
Weekly maintenance includes the following tasks:
•
Deproteinizing the sample path
•
Conditioning the sensors
•
Checking the fill solution levels in the sensors
Deproteinizing the Sample Path
If you are analyzing less than 200 samples per day, deproteinize and condition your
Rapidlab 1200 system weekly. If you are analyzing more than 200 samples per day, or if
your laboratory often processes certain types of blood, high hematocrit samples or blood
with high TPN, you may need to run this procedure more often. Monitor your system and
sensor performance, and adjust the frequency of this procedure as necessary.
You need the following materials for this task:
•
Deproteinizer
•
Syringe
•
Glucose and Lactate test/blank sensors (TB4)
BIOHAZARD:
Wear personal protective equipment. Use universal precautions.
Refer to
Appendix A, Protecting Yourself from Biohazards
for recommended
precautions when working with biohazardous materials.
CAUTION:
Do not expose the glucose and lactate biosensors to bleach,
conditioner, or deproteinizer. Replace the biosensors with the test/blank sensors
(TB4) before conditioning the sensors or deproteinizing the sample path. Reinstall
the biosensors within 2 hours.
CAUTION:
Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the
measurement module frame to discharge static buildup.
Содержание Rapidlab 1200
Страница 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Страница 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Страница 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Страница 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Страница 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Страница 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Страница 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Страница 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Страница 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Страница 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...