
Installation
3 - 9
Siemens AG
SPR2-240.841.01
Page 9 of 10
POWERMOBIL
Medical Solutions
Rev. 01
05.03
CS PS 24
Cables for basic unit
3
•
Insert the cables in the energy chain. Make sure that cables do not intersect.
•
Fasten both cables for the cable module using cable clamps (1/Fig. 21).
•
Close the energy chain.
•
Secure the last link on both ends of the energy chain using cable ties or insulation tape to
avoid accidental opening.
•
Attach the ferrite core to the cable with the plug M2. X1 (1 turn around the ferrite core).
Connect the plug.
•
Attach the connection cable for the inverter. Tighten the nuts with a torque of 4.8 Nm.
Fasten the cable using a cable clamp. Make sure that the shielding makes good ground
contact.
•
Fasten the protective ground wires.
•
Attach the Plexiglas cover (4/Fig. 21).
•
Reconnect the BNC plug for the coaxial cable.
•
Insert the previously ejected contacts of the plug D30 X3 in the connector housing, as
previously noted.
•
Run the cables with their plugs to the corresponding plug connections / PC boards ac-
cording to labeling. Fix the cables using cable ties.
•
Tie the cables close to the lifting column using cable ties in such a way that they do not
rub against any moving parts or are crushed when moving the lifting column.
•
Fix all cables to the cable holder (1/Fig. 21) using cable ties.
•
Reattach all shielded cables using the provided cable clamps. Make sure that the shield-
ing makes good contact (2/Fig. 21) (1/Fig. 22).
Fig. 21
Fig. 22
t