Siemens AG
SPR2-240.831.01.02.02
POWERMOBIL
07.05
CS PS 24
Safety Inspection
35
Page 35 of 44
Medical Solutions
3.5 Measurements
SIE
Protective conductor test (performed under "Final work steps")
•
Test the protective conductor with the protective conductor tester according to
ARTD-002.731.17...
.
•
The protective conductor resistance must not exceed 0.2 Ohms.
NOTE
The protective conductor test is part of the safety inspection. How-
ever, it cannot be performed until all maintenance work has been
completed. See
.
SIE
Leakage current test (performed under "Final work steps")
•
Perform the leakage current test with the system closed according to
ARTD-002.731.17 ...
. Document and evaluate the determined values. Document the
measuring procedure. The leakage current must not exceed 2.5 mA.
NOTE
The leakage current test is part of the safety inspection. However,
it is recommended to perform it only at the end of maintenance,
with the system closed. See
.