7ML19985GG03
Pointek CLS500 – INSTRUCTION MANUAL
Page 17
mmm
mm
P
robe Configur
at
io
n
Process Connection and Seal Configuration of Pointek CLS500
Pressure and Temperature Considerations
The maximum temperature and pressure of operation for the standard Pointek CLS500
level probe is 200
C (392
F) and 200 bar (2900 psi). Please consult the pressure curves
below for qualifications that must be applied to these maximums.
Enamel probes are recommended when the process temperature exceeds 200
C, and/or
in combination with very high pressure.
Process Connection
Seal Type
Seal Description
Threaded
S
Single Cone
Welded Flange
S
Single Cone
Solid Machined Flange
S
Single Cone
HP/HT
Primary graphite seal, and glass seal
Note: Pointek CLS500 HP (high pressure version) is only supplied with enamel
insulation. A primary graphite seal plus a secondary redundant seal is provided
between the electrode and the instrument body.
Note: Consult your local Siemens representative if the material to be measured
may be incompatible with the Pointek CLS500 materials of construction.