Service software
3 - 9
Siemens AG
SPL1-130.840.02
Page 9 of 18
MODULARIS Uro Plus
Medical Solutions
Rev. 03
04.05
CS PS 24
select option: c
3
•
Press the
"c" key
to display the following text::
- *1: sequential number of the card inserted.
- *2: serial number of the card.
- *3: a "gold" card was inserted, i.e. the input slot is closed.
- *4: init-counter in the above example is init=2, i.e. the gold card was logged onto two
different D3 boards.
- *5: n.a.
- *6: n.a.:
- *1:sequential number of the card inserted
- *2: serial number of the card
- *3: a "
p
ay
p
er
u
se" card was inserted
- *4: shock waves still present on the card
- *5: internal system card counter
- *6: number of shock waves registered on inserting the ppu card
Text:
4:
531...0
G
init=2
....
.....
04/13 ...
418
Explan.:
*1
*2
*3
*4
*5
*6
*7
*8
Card ejected
- *7: Date of ejection
- *8: System shock wave counter value when card ejected
Card in system
- *7: Text: "Card locked to unit"
- *8: ............
Text:
5:
531...1
P
89956
0
0
04/13 ... 418
Explan.:
*1
*2
*3
*4
*5
*6
*7
*8
Card ejected
- *7: Date of ejection
- *8: System shock wave counter value when card ejected
Card in system
- *7: Text: "Card locked to unit"
- *8: ............
=> C
Card Data:
# Serial number Type -------< units >-------- eject-date/time pulse-count
GOLD init-cnt ... ... system
PPU on card int. from card
------------------------------------------------------------------------------
1:
2:
3:53148300000000 G init=2 .... .... 04/13/1999 15:48.11 418
4:53149100000001 P 89956 0 0 04/13/1999 16:00.21 418
5:53148300000000 G init=3 .... .... 04/13/1999 16:40.51 418
6:53149100000001 P 84956 0 5000 04/13/1999 17:45.00 418
7:
8:
9:
10:53149100000001 P 84956 5000 0 card locked to unit ......
-- end of card data --