MODULARIS Uro Plus
SPL1-130.840.02
Page 12 of 14
Siemens AG
Rev. 03
04.05
CS PS 24
Medical Solutions
2 - 12
LITHOSTAR MODULARIS Error List
Ultrasound, E90
–
E93
2
E90
Error Text
:
Possible
cause/action:
No connection from LITHOSTAR to SONOLINE
E91
Error Text
:
Possible
cause/action:
Potentiometer skips.
Cable from potentiometer to board D3 ok?
Potentiometer defect?
E92
Error Text
:
The two potentiometers have different values.
Possible
cause/action:
Dampness on the potentiometer.
E93
Error Text
:
Erroneous US values.
Possible
cause/action:
C-MOS has lost values. Perform potentiometer adjustment
(Point "U" in the service menu).
Check default values (see SPL1-130.038.01...).
3
1
2
5
Value between 1 -3 and 2-5
must be identical.
Plug at US holder