LITHOSTAR MODULARIS Error List
High voltage circuit, E60
–
E67
2
LITHOSTAR MODULARIS Circuit diagram J1048 - 12.
E60
Error Text
:
Charging current too low during the charging process.
Possible
cause/action:
Check the line voltage!
Replace the charging / energy unit!
E61
Error Text
:
Overvoltage in the charging unit.
Possible
cause/action:
System OFF
- 10 minute pause -
System ON.
Replace the charging / energy unit!
E62
Error Text
:
Overtemperature in the charging unit.
Possible
cause/action:
Heat in the LITHO MODULARIS ?
System OFF
- 10 minute pause -
System ON.
Replace the charging / energy unit!
E63
Error Text
:
Capacitor defective in the charging / energy unit (out of tole-
rance).
Possible
cause/action:
Replace the charging / energy unit!
E64
Error Text
:
Default voltage not attained.
Current voltage after frequency converter (U
nom
) is greater than
selected voltage.
Possible
cause/action:
Check the fiber optic cable!
1
)
Replace the charging / energy unit!
E65
Error Text
:
Charging voltage too high during the charging process.
Possible
cause/action:
Replace the charging / energy unit!
E66
Error Text
:
Default voltage incorrect.
Possible
cause/action:
Check the fiber optic cable!
1
)
Replace the charging / energy unit!
E67
Error Text
:
No ignition of spark gap after trigger pulse (several times).
Possible
cause/action:
Check the fiber optic cable!
1
)
Replace the charging/ energy unit!