LITHOSTAR MODULARIS Error List
Monitoring, E40
–
E48
2
E40
Error Text
:
Cooling pump fuse defective.
Possible
cause/action:
Fuse F2 on board D3 o.k.?
E41
Error Text
:
Cooling pump M1 defective.
Possible
cause/action:
Motor M1 o.k.?
Circuit diagram J1048 – 13
Connection cable D3.X7 o.k.?
E42
Error Text
:
Coupling pump defective.
Possible
cause/action:
Motor M2 o.k.?
Circuit diagram J1048 – 13
Connection cable D3.X7 o.k.?
E43
Error Text
:
Fuse for coupling pump defective.
Possible
cause/action:
Fuse F3 on board D3 o.k.?
E44
Error Text
:
5 volt power supply defective.
Possible
cause/action:
Replace board D3.
E45
Error Text
:
18 volt power supply defective.
Possible
cause/action:
LED VCC18 on board D3 o.k. ?
Circuit diagram J1048 – 11
Connection cable D3.X8 o.k.?
Check the power supply for the pressure sensors!
E46
Error Text
:
17 volt power supply defective
Possible
cause/action:
LED VCC17 on board D3 o.k. ?
Replace board D3.
E47
Error Text
:
Valve and card terminal fuse defective?
Possible
cause/action:
Short circuit?
Fuse F4 on board D3?
Circuit diagram J1048 – 13
Check valve (coupling circuit).
Check card terminal.
E48
Error Text
:
Reference for temperature sensors defective.
Possible
cause/action:
Replace temperature sensors.
Replace board D3.