5 - 1
Siemens-Elema AB
SPR8-215.898.01
Page 1 of 10
MOBILETT Plus / - E/ - M
Solna, Sweden
Rev. 05
01.02
SPS-UD
Service programs
5
Entering service mode
5
•
Remove the top cover.
To activate the service program, adjust the switch to the service position (S1:1 ->ON, on
board). This will display the program version:
Exit the service mode by adjusting the switch to the operating position (S1:1 -> OFF on
board D1). This will cause program execution to be aborted.
Selecting test programs with the kV+/- buttons
5
There are 13 different test programs available. The display shows the program number in
the kV window, the mAs display will be dark. Example, program 06:
Pressing the "kV+" button will display the next test program. When the last test program is
displayed, pressing the "kV+" button will show the program version again and so on.
Pressing the "kV-" button will display the previous test program. The lamp button has a
toggling function and is used both to start and stop the displayed test program. For exam-
ple, after the lamp button has been pressed, the formation of the capacitor bank starts
and the following is shown on the display:
The test program can be stopped by pressing the lamp button a second time. After the
lamp button is pressed, the display shows:
The program P06 has ended and a new service program can be selected with the kV+/-"
buttons or P06 can be started with the lamp button.
Error messages in service mode
5
Messages appearing in service mode, except in "Fast adaptation" are only displayed, they
are not logged into the memory history.
kV
mAs
vEr x.x = Version x.x (depends on software version installed)
kV
mAs
P06
kV
mAs
CAP For = Capacitor formation program
kV
mAs
P06
StoP