7ML19985CN01
Milltronics SF 500 - INSTRUCTION MANUAL
Page 83
m
mmmm
Communicat
io
ns
Error Codes
The error codes returned in the format area are 8-bit integers found in the lowest 8 bits of
the format word. This allows for 256 potential error codes.
Currently the SF 500 has two error codes available:
Date and Time (R41,000 – 41,006)
The date and time can be read or written in registers 41,000 to 41,006 as defined in the
table above.
Example: If you are located in Toronto, Canada and would like to set the date and time to
February 14, 1999, 1:30 p.m. and 42 seconds, you would write the following:
Values
Description
0
No error
1
Data not available as percent (available as units)
2-255
Reserved
Bits
Values
R41,000
1999
R41,001
2
R41,002
14
R41,003
13
R41,004
30
R41,005
42
R41,006
-5
Note:
The time zone register is used only as a reference and does not affect the
operation of the SF 500.