Page 36
Milltronics SF 500 – INSTRUCTION MANUAL
7ML19985CN01
mmmm
m
Recalibr
at
ion
Routine Span
Press
if
Press
if
This indicates that the mechanical system is errant. Use P388, initial span, judiciously and
only after a thorough mechanical investigation.
Find and rectify the cause of the increased or decreased deviation. Then re-try a span
recalibration.
If this deviation is still unacceptable, set P388 to 1 to invoke an initial span calibration.
Further deviation limits are now based on this new initial span.
Press
This is the end of span calibration. Remove the test weight and return to RUN.
Note:
To obtain an accurate and successful calibration, ensure that no material is
flowing through flowmeter and the test weight is applied
Span Calibration. Current Span
41285
Setup test. Press ENTER to start
Zero should be done prior to Span
Setup test. Press ENTER to start.
Span Calibration in progress
Current Reading:
55.56 t/h
Calibration complete. Deviation
0.03
Press ENTER to accept value
41440
Span Count too Low.
Press CLEAR to continue.
Calibration is out of range
Deviation Error:
Span Calibration. Current Span
41440
Setup test. Press ENTER to start
e.g. the current span count
do a zero calibration or press
the rate reported while calibration is
in progress.
e.g. the deviation from the previous
span
e.g. the new span count, if accepted
if the signal from the load cell or
LVDT is too low, ensure proper test
weight is applied during span
check for proper load cell /LVDT wiring
if unacceptable, press
to restart
e.g. span calibration is accepted and
displayed as the current value