Error Codes
Error
Code
Code Name
Message/Action
201
Error - Load Cell A & B
Reading between A & B > 20000, or no signal. Check
wiring.
203
Err: 203
Memory failure test. Consult Siemens Milltronics.
204
Integrator not configured
P002-P017 must be programmed.
205
Err: 205
Zero or span calibration required.
210
Remote Totalizer 1
exceeded
Increase resolution.
211
Remote Totalizer 2
exceeded
Increase resolution.
213
Maximum rate exceeded
Rate is > thee times the design rate. If no mechanical
cause, check to see if re-rating the design rate is
required.
220
Span too low
Span is < 1 mV. Insure proper test weight or chain is
applied during span.
221
Span out of range
Span deviation > 12.5%. Consider an initial zero (P377).
Refer to
Initial Zero
on page 34.
222
Zero out of range
Zero deviation > minimum limit. Consider an initial Span
(P388). Refer to
Initial Span
on page 37.
223
Security Violation
An attempt to run command / calibration that is not
allowed under current security level.
224
Function not permitted
Function not allowed under current security level.
225
BF
Flashes in the bottom right corner of display when bat-
tery charge is too low.
226
Load Cell AD’s are not func-
tioning
Consult Siemens Milltronics.
227
Err: 227
No process data available. Consult Siemens Milltronics.
228
Batch pre-act adjustment
> 10%
Pre-act adjustment is ignored. Tune process to limit batch
error.
240
Integrator not configured
P002-P017 must be programmed
241
No PID mA Input
PID Process Value Source (P402) or PID Setpoint (P414)
has been programmed for a mA Input, however mA Input
Function (P255) has not been programmed properly.
242
No PID mA Output
PID System (P400) has been turned on, but mA Output
(P201) has not been programmed properly.
243
No batch setpoint relay
Batch has been set up, but no relay has been configured
for a setpoint.
PF
Power Failure
Displayed at the bottom right corner of the display if
power is interrupted after the integrator has been
calibrated.