
7ML19985CN01
Milltronics SF 500 - INSTRUCTION MANUAL
Page 105
mm
mmm
Pa
ra
m
ete
rs
P367 Direct Zero Entry
Directly enters the zero reference count.
Direct entry is intended for use when replacing software or hardware and it is not
convenient to perform an initial zero at that time. (
f
= 0)
Refer to
Recalibration
on page 30 for execution.
P368 Direct Span Entry
Directly enters the span reference count for the span selected, span 1 to 8 (P368-01 to
-08).
Direct entry is intended for use when replacing software or hardware and it is not
convenient to perform an initial zero at that time. (
f
= 0)
Refer to
Recalibration
on page 30 for execution.
P370 Zero Limit Deviation %
Sets the zero calibration deviation limit (
±
) from the last initial zero. If the accumulated
deviation of successive zero calibrations exceeds the limit, the zero calibration is aborted.
(
f
= 12.5)
Enter the maximum allowable % deviation.
P377 Initial Zero
Resets the initial zero.
The initial zero is the reference zero to which all subsequent operator initiated zero
calibrations are compared in determining whether they have deviated beyond the zero
limit (P370). (
f
= 1)
P388 Initial Span
Resets the initial span for the span selected, multispan 1 to 8 (P388-01 to –08).
The initial span is the reference to which all subsequent span calibrations are compared
in determining whether they have deviated beyond an accumulated
±
12.5% of the initial
span. (
f
= 1)
Note:
Refer to
Initial Zero
on page 34 for execution.
Note: Refer to
Initial Span
on page 37 for execution.
Содержание Milltronics SF 500
Страница 1: ...Instruction Manual September 2003 SF 500 milltronics ...
Страница 6: ...iv mmmmm Table of Contents ...