
Page 94
Milltronics SF 500 - INSTRUCTION MANUAL
7ML19985CN01
mmmm
m
Pa
ra
m
et
er
s
P008 Date
Enter the current date in
yyyy-mm-dd
format.
Where:
yyyy = year
mm = month, 01 –12
dd = day, 01 – 31
e.g. 1999-03-19 (March 19, 1999)
P009 Time
Enter the current time in
hh-mm-ss
, 24 hour format.
Where:
hh = hour
mm = minute
ss = second
P011 Design Rate
Specifies the design rate of material flow for the flowmeter. (
f
= 0.00)
Enter the design rate from the supplied design data sheet, in the units selected (P005).
P017 Test Load
The rate referenced when performing a span. (
f
= 0.00)
Enter the test rate value as shown in the corresponding solid flowmeter instruction
manual.
P019 Manual Span Adjust
Provides a means for adjustment to the span calibration. (
f
= 0)
The adjustment value is determined by performing material tests and is subsequently
entered either as a calculation of % change into P598, or as the weight of the material
test.
Entry:
1 = % change
2 = material test
Refer to
Recalibration
on page 30.
Содержание Milltronics SF 500
Страница 1: ...Instruction Manual September 2003 SF 500 milltronics ...
Страница 6: ...iv mmmmm Table of Contents ...