Description
3.6 Functions
MicroSAM
Operating Instructions, 06/2012, C79000-G5376-C560-07
53
3.6.9
Temperature control systems
Previously available temperature control systems
The device previously featured three temperature control systems:
● Two temperature control systems heated the column and the detectors. They are located
in the analysis module.
● The third temperature control system heated the pneumatic connectors.
Additional fourth temperature control system
The fourth temperature control system is located on devices as of release version 10 at the
sample inlet of the MicroSAM:
①
Fourth temperature control system at the sample inlet
②
Sample path in the heated copper tube
③
Access to the pneumatic connector