Hardware maintenance

PM

Fans - cleaning

Clean dust from all fans.

PM

Check internal cables and plugs

Check that all internal cables are properly connected to the devices.

PM

Cleaning

If necessary, clean dust from the computer cabinet.

Reinstall the computer access panel and connect the power supply cable.

2.2

Cables

PM

Check that cables are not squeezed or broken

Check that the cables are neither squeezed nor broken; if they are, they must be 
replaced.

PM

Check all external cable connections

Check that all external cables are properly connected.

PM

Safe cable run

Check that all cables are placed in such a way that no one will trip over them.

2.3

Keyboard and mouse

PM

Keyboard - cleaning

Use a damp cloth to clean dust from the keyboard.

PM

Mouse - operational check

Check that the mouse is operational. 

PM

Mouse - cleaning

Open the mouse and clean the rolls inside the housing.

Clean the external surface of the mouse.

2.4

Monitors

PM

Desktop placement

Ensure that the two high resolution monitors are safely placed on the workstation table.

Ensure that the service monitor is safely placed on a side table.

PM

Cleaning

Clean the housing of the two high resolution monitors and the service monitor with a 
damp cloth.

Wipe the glass with a cloth moistened with a lukewarm aqueous solution of mild 
detergent. 

Содержание MammoReport Plus

Страница 1: ...contents is not permitted without express written authority Offenders will be liable for damages All rights including rights created by patent grant or registration of a utility model _or_ design _are_ reserved The Maintenance Protocol SPB7 420 832 20 01 02 is required for these Instructions SP System Manual Maintenance Instructions System maintenance ...

Страница 2: ...iemens or one of its affiliates to provide such services Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro cedures Copyright notice The DICOM functionalities in the MammoReportPl...

Страница 3: ...ation 1 1 Training 1 1 Documents required 1 1 Time required 1 1 Text emphasis 1 2 Explanation of abbreviations 1 2 2 _______Hardware maintenance __________________________________________ 2 1 Computer 2 1 Cables 2 2 Keyboard and mouse 2 2 Monitors 2 2 3 _______Check of image quality __________________________________________ 3 1 4 _______Changes to previous version ________________________________...

Страница 4: ...0 2 Contents MammoReport Plus SPB7 420 831 20 Page 2 of 2 Siemens Medical Solutions Rev 01 01 05 CSPS 24 Erlangen Germany Page ...

Страница 5: ...ammoReportPlus Maintenance Protocol SPB7 420 832 20 1 1 Safety Information The product specific safety information contained in this document as well as the general safety information must be observed The general ESD guidelines must be observed Refer to document ARTD 001 709 01 1 2 Training The training required is described in the document Field Service Strategy SPB7 420 890 10 1 3 Documents requ...

Страница 6: ...e to property NOTICE used without the safety alert symbol indicates a risk of danger that if disregarded leads or may lead to a potential situa tion which may result in an undesirable result or state other than death physical injury or damage to property NOTE contains information provided with special emphasis to facilitate proper use of the equipment or proper execution of a procedure i e hints t...

Страница 7: ...ld flow from front to back Listen to the noise of the fans Does it sound ok Shut down the operating system properly and turn off the computer and any external devices Disconnect the power supply cable High Voltage Life threatening electric shock hazard exists Do not touch computer components until the power supply cable has been disconnected The computer contains electrostatic highly sensitive com...

Страница 8: ...ions Check that all external cables are properly connected PM Safe cable run Check that all cables are placed in such a way that no one will trip over them 2 3 Keyboard and mouse PM Keyboard cleaning Use a damp cloth to clean dust from the keyboard PM Mouse operational check Check that the mouse is operational PM Mouse cleaning Open the mouse and clean the rolls inside the housing Clean the extern...

Страница 9: ...5 CSPS 24 3 Check of image quality QIQ Calibration of High resolution monitors The two High resolution monitors must be calibrated every year Perform calibration according to the MammoReportPlus System Manual Calibration of Monitors SPB7 420 820 20 Check calibration according to the MammoReportPlus Quality Control Manual SPB7 420 621 20 ...

Страница 10: ...MammoReport Plus Register 8 SPB7 420 831 20 Page 2 of 2 Siemens Medical Solutions Rev 01 01 05 CSPS 24 Erlangen Germany 3 2 Check of image quality This page intentionally left blank ...

Страница 11: ...to previous version 4 1 Siemens Medical Solutions Register 8 SPB7 420 831 20 Page 1 of 2 MammoReport Plus Erlangen Germany Rev 01 01 05 CSPS 24 4 Changes to previous version Chapter Page Change Initial version ...

Страница 12: ...MammoReport Plus Register 8 SPB7 420 831 20 Page 2 of 2 Siemens Medical Solutions Rev 01 01 05 CSPS 24 Erlangen Germany 4 2 Changes to previous version This page intentionally left blank ...

Отзывы: