Siemens AG
SPB7-250.815.06.02.02
MAMMOMAT Novation DR
12.05
CS SD 24
Prerequisites
7
Page 7 of 116
Medical Solutions
Meters, appliances and tools
0
Auxiliary Materials
0
NOTE
Calibrated instruments are required.
Tab. 1
Meters, appliances and tools
Phantoms / Auxiliary
0
Item
Remarks
Material
Number
With
Sys-
tem
Oscilloscope >50MHz with memory e.g. TEKTRONIK 314
No
Digital multimeter including an mAs
meter
e.g. FLUKE 8060A or
FLUKE 87
97 02 101
97 03 976
No
Service PC
See the CS Intranet
(
CS/For Service/Com-
mon Services/Service
Laptop for CSEs
) for
details.
No
Power line impedance meter
84 28 104
No
Power ground-wire tester
44 15 899
No
Luminance meter (for monitor cali-
bration)
e.g. Mavo Monitor
or Wellhöfer LXplus (05
146 167)
97 02 432
No
Luminance meter (measures the
light intensity from the X-ray field)
e.g. SMfit Mammo
88 81 281
No
Dose meter
e.g. Solidose
88 81 323
No
Ion chamber
For dose meter Solidose 88 81 315
No
Densitometer
e.g. X-Rite 331
97 02 416
No
A non-invasive digital kV meter
If not available in the
district, a scope can be
used instead
No