Power up the MAMMOMAT
21
Page 21 of 116
Medical Solutions
Measuring the line resistance
0
1. Connect the line resistance meter to the mains input terminal on the stand.
2. Mains supply ON.
3. Carry out measurement.
4. Mains supply OFF.
To achieve the maximum output, the line resistance must not exceed the following values
Tab. 5
Measuring the line resistance
Start-up of the MAMMOMAT system
0
Switching on the system
1. Switch
ON
the external main power switch in the room.
2. Verify that the generator and BRICK power up properly.
3. Wait
5 minutes
.
4. Press the power
ON
button on the control console to activate the MAMMOMAT Nova-
tion DR. The internal monitoring system automatically performs a functional check of
the MAMMOMAT Novation DR. As a result dr is displayed on the film density display on
the control panel to indicate that the communication with the MAMMOMAT Novation
DR system is working properly when the detector wing is selected.
5. Switch
ON
the PC and the screen at the acquisition workstation.
Checking the Brick and Acquisition Workstation communication
Verify that the communication between the BRICK and the acquisition workstation is
working properly (LEDs V2 and V3 on the BRICK must be green). The PXCM includes
three status LEDs:
Nominal Input Voltage
[VAC]
Line Resistance Ri
max
[Ohm]
208
0.45
230
0.50
240
0.60
277
0.65
400 (2-phase)
0.85